Performance Specifications


FLOW ACCURACY

± 1% of reading ± 0.2% of full scale*

*Point velocity for insertion flowmeters. T.F. Hudgins recommends a minimum of 15 diameters of straight pipe upstream of the flowmeter and 10 diameters downstream for insertion flowmeters. T.F. Hudgins recommends a minimum of 8 diameters of straight pipe upstream of the flowmeter and 4 diameters downstream for inline flowmeters.


FLOW REPEATABILITY

± 0.2% of full scale


FLOW RESPONSE TIME

0.9 seconds (one time constant)


TEMPERATURE ACCURACY

± 1.8° F (± 1.0° C) over -40 to 250° F (-40 to 121° C)

± 3.6°F (± 2.0° C) over 250 to 650° F (121 to 343° C)

Minimum velocity 60 SFPM