TFH 3000 Performance Specifications

FLOW ACCURACY: INLINE METER

± 1% of reading ± 0.2% of full scale*

*T.F. Hudgins recommends 8 diameters of straight, unobstructed pipe upstream of the flowmeter and 4 diameters downstream.


FLOW ACCURACY: INSERTION METER

± 1% of reading ± 0.4% of full scale*

*T.F. Hudgins recommends 15 diameters of straight, unobstructed pipe upstream of the flowmeter and 10 diameters downstream.


FLOW REPEATABILITY

± 0.2% of full scale


FLOW RESPONSE TIME

0.9 seconds (one time constant)


TEMPERATURE ACCURACY

± 1.8° F (± 1.0° C)  -40 to 250° F (-40 to 121° C)

± 3.6°F (± 2.0° C), 250 to 650° F (121 to 343° C)

60 SFPM Minimum


CALIBRATION

Factory calibration to NIST traceable standards

 


CAL-V and Zero CAL-CHECK

In situ, operator-initiated calibration validation